Is a Peristaltic Pump Self-Priming? Suction, Air, Start-Up?

Yes, a peristaltic pump is self-priming. The squeezed tube pulls liquid in by itself. That is peristaltic pump self-priming. You do not fill the pump first.

In this guide, you will learn about suction, air, dry running and start-up. With its SC-40 hose pump, Innofluid is able to lift liquids up to 9.5 meters. A term for this is a suction lift. At this location, there is only one lift number. This guide fills up the gaps left by the site for other models.

What Does a Self-Priming Pump Do?

A self priming pump pulls liquid in on its own. It does not need liquid inside the casing first. The casing is the outer shell of the pump. Suction lift is how high a pump can pull liquid up. It is measured in metres.

Some pumps need liquid in the casing before they start. Without it, they only move air. The priming function takes away that extra step. Peristaltic pump self-priming is part of how the tube works.

How Peristaltic Pump Self-Priming Works Inside the Tube

Peristaltic pump self-priming comes from the tube. A motor turns a rotor, which is the spinning part. The rotor has rollers. The rollers press on a soft tube.

Pinch, Push and Refill

  1. Pinch: a roller squeezes the tube flat.

  2. Push: a small pocket of liquid moves ahead of the roller.

  3. Refill: the tube goes back to its round shape. It pulls in more liquid.

The tube makes a vacuum as it springs back. The vacuum pulls new liquid in from the inlet. At least one roller always pinches the tube. So the pump works as its own valve. Liquid cannot slip back. Only the tube touches the liquid.

More rollers make smaller pockets of liquid. This gives a smoother flow. The EasyPump head has 4 rollers. This means low pulsation (less pulsing) and high precision.

Same Volume Every Turn

A peristaltic pump is a positive displacement pump. It moves the same volume of liquid on each turn. On the LabV6-III, the 18# tube moves 4.3 mL per turn. At 600 rpm (turns per minute), it reaches 2580 mL/min.

What Happens With an Empty Inlet Line?

The inlet line is the hose that brings liquid to the pump. The site says you do not need to fill the casing first. The SC-40 pulls liquid from a lower tank with no pre-fill.

The site gives no empty-line figure for lab pumps. So test peristaltic pump self-priming on a dry hose with your own set-up. You can also ask about your model on the contact page: innofluid.com/contactus.html.

How High Can the Pump Pull Liquid Up?

The SC-40 hose pump pulls liquid up to 9.5 m. That is the published lift. It matters most when you pull liquid up an empty inlet line.

SC-40 Hose Pump Figures

Item

Published figure

SC-40 hose pump: suction lift

up to 9.5 m

SC-40 hose pump: self-priming

yes

SC-40 hose pump: check valves

none needed

These figures are only for the SC series. They are not for the lab pumps. The lab pump pages show speed and flow, but no lift.

Can It Handle Air, Gas and Running Dry?

The SC-40 page lists gaseous liquids. It also says the pump can run dry with no time limit and no damage. These claims are on the SC page only.

Ask about your own model before you rely on them, even with a self priming pump. The site makes no air or dry-run claim for its lab pumps. So do not assume one.

Good air handling matters most when the inlet line starts empty. Air can hurt other pumps too. In a centrifugal pump, air and cavitation can cause trouble. Cavitation is when tiny bubbles form and burst. This can damage parts.

How Do Centrifugal and Diaphragm Pumps Start Up?

Unlike a self priming pump, other pumps start in different ways.

Start-Up by Pump Type

Pump type

How it starts

Peristaltic

Self-priming

Centrifugal

Usually needs liquid in the casing. Running dry can damage the seal.

Diaphragm

Priming depends on the valves.

A check valve is a one-way flap. It stops liquid from flowing back. A peristaltic pump has no valve seat where solids can get stuck. That is one reason peristaltic pump self-priming stays simple.

How Do You Set Up a Lab Pump the First Time?

  1. Pick a tube that fits the pump head. The KD15/KD25 head has a flip cover. It opens the pressure block, so you can change tubes fast.

  2. Check the tube's pressure rating. Wider tubes take less pressure.

  3. Make sure the tube suits your liquid. Innofluid lists silicone, PharMed BPT, several Tygon grades, Chem-Durance, Versilon F-5500-A and Viton tubing.

  4. Set the speed or working mode. The LabV6-III runs from 0.1 to 600 rpm. It has four modes: fixed volume, volume dispensing, speed dispensing and timing start/stop.

  5. Calibrate after every tube change. This means you check and fix the flow. The LabV6-III has intelligent calibration. You can also adjust the flow while the pump runs.

Priming Function, Back Suction and Safety Stop

Priming function. The site has no separate priming mode. The LabM3 has a full speed button. It fills and empties the line. This is from its product page.

Back suction. This is a setting from 0 to 360°. It pulls a little liquid back when a dose ends. It does not pull liquid in at the inlet. Safety stop. This is an optional stop for the KD15/KD25 head. It works when the lid opens. The LabV6-III page has video tutorials. Watch the Basic Setup Demonstration first.

How Do Tube Size and Gap Change the Flow?

Tube size sets how much liquid moves in each turn. It also sets the pressure limit. These figures are from the LabV6-III page.

LabV6-III Tube Size, Flow and Pressure

Tube

Size (ID × wall)

mL per turn

Flow at 0.1–600 rpm

Max pressure (intermittent / continuous)

13#

0.8 × 1.6 mm

0.053

0.0053–32 mL/min

0.27 / 0.17 MPa

25#

4.8 × 1.6 mm

1.967

0.1967–1180 mL/min

0.24 / 0.14 MPa

18#

7.9 × 1.6 mm

4.3

0.43–2580 mL/min

0.10 / 0.07 MPa

The gap between the roller and the track is called occlusion. If the gap is loose, liquid slips back. If it is tight, the tube wears out faster. EasyPump I, II and V have a fixed gap. III, IV and VI have a gap you can adjust.

Who Makes These Pumps?

Innofluid is based in Lingang New Area, Shanghai. Its home page lists 119 patents. It says the company serves 110 countries. 26% of the team work in research and development.

The LabV6-III has a 3-year warranty. Its flow is accurate to within ±0.5%. The company also offers OEM pumps, pump heads and ODM service.

FAQs

Do peristaltic pumps need priming?

No, a peristaltic pump does not need priming. It is self-priming. The squeezed tube makes a vacuum and pulls the liquid in. You do not fill the pump first. The SC-40 has a suction lift of up to 9.5 m. For other models, ask for the number.

How high can a peristaltic pump lift liquid?

The SC-40 hose pump can lift liquid up to 9.5 m. That is the only lift number on the Innofluid site. The lab pump pages, like the LabV6-III and LabM3, show flow and speed only. They show no lift. For your own model, ask for the number first.

Can a peristaltic pump run dry?

The SC-40 page says its pump can run dry with no time limit and no damage. This is for the SC line only. The site does not say this for its lab pumps. Check your own model first. Send the model name on the contact page before you test.

Can a peristaltic pump handle air or gas?

The SC-40 page lists gaseous liquids, so that pump can handle them. The site says nothing about air or gas for its lab pumps. Air can harm other pumps, like centrifugal pumps. For a lab pump, ask about air handling before you run a gassy liquid.

What does back suction do?

Back suction pulls a little liquid back at the end of a dose. The setting is 0 to 360° on the LabV6-III and LabM3. It helps stop drips at the outlet. It does not pull liquid in at the inlet. So it is not a priming setting.

Does a peristaltic pump need check valves?

No, a peristaltic pump does not need check valves. At least one roller always pinches the tube, so liquid cannot flow back. The SC-40 page also says no check valves are needed. A diaphragm pump is different. Its priming depends on the valves.

Conclusion

Peristaltic pump self-priming means you do not fill the pump first. The squeezed tube pulls the liquid in. The SC-40 can lift liquid up to 9.5 m. For a lab pump, check the tube and the speed before the first run.

Innofluid Co., Ltd. shows the tube sizes, flow and pressure for its lab pumps. This makes it easy to pick the right pump. Send your flow, tube size and liquid to info@innofluid.com. Then ask about your model.

 
logo
About Us

Company Profile
Product

Laboratory Peristaltic Pump
Industrial Peristaltic Pump
Explosion proof peristaltic pump
Dispensing and Filling Peristaltic Pump
Multi-Channel Independent Control System
ODM Fluid Handling Solution
OEM Pump and Pump Head
Tubing and Accessories
Liquid Filling Machine
Syringe Pump Series
Application

Laboratory application
Industry application
E Liquid Filling Pump
Filling Various Reagents
Support

Download

Contact

Innofluid Co., Ltd.
Add: Building 10, No. 860, Xinyang Road, Lingang New Area, Pilot Free Trade Zone, Shanghai, China
Zip Code:200135
Tel: +86-312-5958380
Website: http://www.innofluid.com
E-mail: info@innofluid.com
    Contact
    © 2023 申辰流体科技(上海)有限公司 | 沪ICP备2022015939号-2 | [Privacy Policy]

    We use cookies

    Our website uses cookies to enhance user experience, personalise content and show you more relevant adverts about our great products, also to analyze our website traffic.